SC4500 Controller, Modbus TCP + LAN, 1 digital Sensor, 100-240 VAC, without power cord

SC4500 Controller, Modbus TCP + LAN, 1 digital Sensor, 100-240 VAC, without power cord
Product #: LXV525.99A05501
Unit Price お問い合わせ

Easy Adoption

The familiar experience of a modern touchscreen, the ability to use your current Hach sensors, and the same footprint as the SC200, make installation and integration of the SC4500 Controller seamless.

No Time for Downtime

The SC4500's built-in predictive diagnostic software ensures measurement confidence and reduces the risk of unexpected equipment downtime by enabling proactive maintenance planning via MSM, including step-by-step instructions.

The Connectivity Options You Need

The Controller provides local communication to SCADA or a PLC, as well as remote access through a secure, cloud-based connectivity option to integrate with Claros, the Water Intelligence System from Hach. From analog and advanced digital protocols to Wi-Fi, cellular or LAN, the SC4500 gives you the flexibility to adapt in a rapidly changing world.

What's in the box?

SC4500 Controller, Non-Claros, without power cord; includes mounting hardware

仕様

Analogue Output Functional Mode: Linear, PID
Communication (optional): Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.

Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Compatible network technologies: GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.)
CDMA (e.g. Verizon)
Conduit Openings: ½" NPT conduit
Dimensions: ½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)
Indoor/Outdoor: 直射日光や紫外線の当たる屋外に設置する場合は、紫外線防止スクリーンやサンルーフが必要です。
Installation category: Category II
Material Enclosures: Polycarbonate, aluminum (powder coated), stainless steel
Network connectivity: LAN: Two Ethernet connectors (10/100 Mbps)
Cellular: External 4G
Wi-Fi
Sensor Input #1: Digital
Sensor Input #2: None
Software available: Non-Claros
USBポート: Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.
コンプライアンス認証: CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco)
リレーEU: Two relays (SPDT);
Wire gauge: 0.75 to 1.5 mm² (18 to 16 AWG)

AC controller
Maximum switching voltage: 100 - 240 VAC
Maximum switching current: 5 A Resistive/1 A Pilot Duty
Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty

DC controller
Maximum switching voltage: 30 VAC or 42 VDC
Maximum switching current: 4 A Resistive/1 A Pilot Duty
Maximum switching power: 125 W Resistive/28 W Pilot Duty
汚染度: 4
高度: 3000 m maximum
質量: 1.7 kg (controller only, w/o modules)
取り付け方法: Wall, Pole, or Panel Mounting
出力: Modbus TCP & LAN
所要電力 (ボルト): 100-240 VAC ±10%, 50/60 Hz; 1 A
製品に含まれるもの:: SC4500 Controller, Non-Claros, without power cord; includes mounting hardware
説明: Microprocessor-controlled and menu-driven controller that operates the sensor
測定: Two device digital SC connectors
対応機種: Compatible Sensors and Analysers / Software Version (Release Year)
Amtax sc / V2.30 (2018) or higher
A-ISE sc / V1.02 or higher
AN-ISE sc / V1.08 (2013) or higher
N-ISE sc / V1.02 or higher
Nitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higher
NT3100sc/NT3200sc
Phosphax sc / V2.30 (2018) or higher
Phosphax sc LR/MR/HR / V1.01 (2018) or higher
TSS sc / V41.73 (2013) or higher
Solitax sc / V2.20 (2013) or higher
TU5300sc, TU5400sc / V1.34 (2017) or higher
SS7 sc (in Bypass) / V1.01 (2006) or higher
Ultraturb sc / V3.06 (2017) or higher
1720E / V2.10 (2006) or higher
Sonatax sc / V1.15 (2016) or higher
CL17sc / V2.7 (2019) or higher
CL10sc / V1.14 (2013) or higher
9184sc, 9185sc, 9187sc* / V2.03 (2013) or higher
Uvas plus sc / V3.01 (2017) or higher
LDO 2 sc* / V1.22 (2013) or higher
3798sc* / V2.03 (2013) or higher
3700sc + Inductive Conductive Digital 6120800 / V3.00 (2017) or higher
3422sc + Contacting Conductive Digital 6120700 / V3.00 or higher
3700 analog + Conductivity Module LXZ525.99.D0004
3400 analog + Conductivity Module LXZ525.99.D0004
pHD sc*, pHD-S sc / V3.10 (2016) or higher
1200-S sc* / V2.04 (2013) or higher
pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higher
pHD analog + pH/ORP Module LXZ525.99.D0003
RC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher
RC and PC analog + pH/ORP Module LXZ525.99.D0003
8362sc* / V3.00 (2017) or higher
Polymetron pH/ORP analog + Ultrapure pH/ORP Module LXZ525.99.D0007
Polymetron Conductivity analog + Ultrapure Conductivity Module LXZ525.99.D0006
GS1440 and GS2440EX Sensors H₂S
FP360 sc / V1 or higher

*Hardware Version1 of instrument is not supported
電源: Without power cord
動作温度範囲: -20 to 60 °C (-4 to 140 °F) (8 W (AC)/9 W (DC) sensor load)
-20 to 45 °C (-4 to 113 °F) (28 W (AC)/20 W (DC) sensor load)
Linear derating between 45 and 60 °C (-1.33 W/°C)
表示部: 3.5-inch TFT colour display with capacitive touchpad
保管条件: -20 to 70 °C, 0 - 95% relative humidity, non-condensing
保護クラス: I, connected to protective earth
保証: 12 ヶ月
防水構造: UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X
Metal enclosure with a corrosion-resistant finish