Automatic cleaning:
Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Demineralised water:
Internal dilution option: Rinse water and Dilution water
Digital outputs:
Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
RS485 communication: Profibus DP or Modbus RTU
Drain:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Measurement method:
Discontinuous, direct measurement by combined Ion-Selective Electrode with standard addition
Number of sample streams:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Reagent requirements:
Keep between 10 - 30 °C (50 - 86 °F)
Sample Quality:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Validation:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
アナログ出力:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
サンプル圧力:
By external overflow vessel (open to atmospheric pressure)
サンプル温度:
10 - 30 °C (50 - 86 °F)
サンプル流量:
100 - 300 mL/min
検出下限(LOD):
No dilution
2 - 100 mg/L Cl-: 2 mg/L
5 - 250 mg/L Cl-: 5 mg/L
10 - 500 mg/L Cl-: 10 mg/L
20 - 1000 mg/L Cl-: 20 mg/L
Internal dilution
100 - 5000 mg/L Cl-: 100 mg/L
200 - 10000 mg/L Cl-: 200 mg/L
原材料:
Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
校正:
Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
質量:
Max. 35 kg (77 lb)
周囲温度:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
寸法:
688 mm x 460 mm x 340 mm
精度:
Better than 2% full scale range for standard test solutions
測定項目:
Chloride
測定周期:
No dilution
Default: 5 minutes
Continuous: 4 minutes
Internal dilution
Default: 10 minutes
Continuous: 7 minutes
測定範囲:
No dilution
2 - 100 mg/L Cl-
5 - 250 mg/L Cl-
10 - 500 mg/L Cl-
20 - 1000 mg/L Cl-
Internal dilution
100 - 5000 mg/L Cl-
200 - 10000 mg/L Cl-
測定方法:
Ion-Selective Electrode with standard addition
電源:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
認証:
CE, ETL certified to UL and CSA safety standards, UKCA
保護クラス:
IP44
保証:
12 ヶ月
妨害物質:
Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.