Automatic cleaning:
Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Demineralised water:
Internal dilution option: Rinse water and Dilution water
Digital outputs:
Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
RS485 communication: Profibus DP or Modbus RTU
Drain:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Measurement method:
Discontinuous measurement by combined Ion-Selective Electrode with standard addition
Number of sample streams:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Reagent requirements:
Keep between 10 - 30 °C (50 - 86 °F)
Sample Quality:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Validation:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
アナログ出力:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
サンプル圧力:
By external overflow vessel (open to atmospheric pressure)
サンプル温度:
10 - 30 °C (50 - 86 °F)
サンプル流量:
100 - 300 mL/min
検出下限(LOD):
No dilution
0.2 - 10 mg/L F-: 0.2 mg/L
0.5 - 25 mg/L F-: 0.5 mg/L
1 - 50 mg/L F-: 1 mg/L
2 - 100 mg/L F-: 2 mg/L
Internal dilution
10 - 500 mg/L F-: 10 mg/L
20 - 1000 mg/L F-: 20 mg/L
原材料:
Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
校正:
Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
質量:
Max. 35 kg (77 lb)
周囲温度:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
寸法:
688 mm x 460 mm x 340 mm
精度:
Better than 2% full scale range for standard test solutions
測定項目:
Fluoride
測定周期:
No dilution
Default: 5 minutes
Continuous: 5 minutes
Internal dilution
Default: 10 minutes
Continuous: 8 minutes
測定範囲:
No dilution
0.2 - 10 mg/L F-
0.5 - 25 mg/L F-
1 - 50 mg/L F-
2 - 100 mg/L F-
Internal dilution
10 - 500 mg/L F-
20 - 1000 mg/L F-
測定方法:
Ion-Selective Electrode with standard addition
電源:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
認証:
CE, ETL certified to UL and CSA safety standards, UKCA
保護クラス:
IP44
保証:
12 ヶ月
妨害物質:
Metal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.