Orbisphere Hydrogen (H₂) TC Sensor with nitrogen purge, protection cap, ext. temp. sensor (20 bar)

Orbisphere Hydrogen (H₂) TC Sensor with nitrogen purge, protection cap, ext. temp. sensor (20 bar)
Product #: 31290TC
Unit Price お問い合わせ

Unique thermal conductivity (TC) sensor design

Selective and precise measurement of hydrogen in gaseous or liquid samples

No specific sample preparation required, the gas concentration is measured directly in the sample

Rugged construction can handle harsh plant conditions and high sample pressures

Flexibility of installation for better process control and very low maintenance

What's in the box?

Sensor only.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.

仕様

Purge connection: Nitrogen Purge
アプリケーション: Power
サンプル圧力: 0 - 20 bar
サンプル温度: 0 - 50°C (32 - 122°F)
温度範囲: -5 - 50 °C
質量: 0.95 kg
精度: (Sample temp. 0 - 50°C independent of calibration temperature)
The greater of
Membrane 2952A: ±3% of reading, or ±60 ppb, or ±0.6 cc/kg, or ±20 mbar
Membrane 2935A: ±3% of reading, or ±150 ppb, or ±2.5 cc/kg, or ±50 mbar
  (Sample temp. 20 - 50°C within ± 5°C of calibration temperature)
The greater of
Membrane 29561A: ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar
Membrane 2952A: ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar
Membrane 2935A: ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar
製品に含まれるもの:: Sensor only.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.
測定周期: 17 s
耐放射線性レベル: Membrane 29561A: 105 rad
Membrane 2952A: 108 rad
Membrane 2935A: 108 rad
認証: 2004/108/EC - EN 61326-1
反応時間: 17 s
範囲: Range at 25 °C
Membrane 29561A: 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar
Membrane 2952A: 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar
Membrane 2935A: 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar
保護クラス: IP65
保証: 6 ヶ月
流量: Typical Flow Rate
Membrane 29561A: 220 mL/min
Membrane 2952A: 200 mL/min
Membrane 2935A: 100 mL/min